| Catalog Number |
05.001.240 |
| Medical Device Problem Code |
Intermittent Loss of Power (4016)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/07/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H10 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
|
| |
|
Event or Problem Description
|
|
It was reported that the sagittal saw device was freezing randomly.It was not reported if the device was used in surgery.It was not reported if there were any delays in a surgical procedure or if a spare device was available.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: investigation summary.D9: the actual device was returned for evaluation.H3, h6: a visual and functional assessment was performed on the device the following steps failed: leakage test using bubble emission technique.Check general function of device.Check oscillation frequency with frequency meter.During the service evaluation the following failures were identified: internal finding : seized bearing.Internal finding : leak tightness test failure.Functional : will not run.Conclusion: failure reported is not confirmed.Root cause: improper maintenance (3214) action: exchange, replacement.Device history review.No manufacturing record evaluation required as no manufacturer or service related issue found.
|
| |
|
Search Alerts/Recalls
|