(b)(4) date sent: 4/13/2026 d4: batch #unk d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.D4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.H4: the device manufacture date is currently not available.Additional information was requested, and the following was obtained: is the current patient status known? unknown.Was there any patient consequence or change in the post-operative care of the patient as a result of the event? unknown.Was the firing sequence started but not completed? if yes: yes, the surgeon said it sounded like the battery died.Did the device deliver any staples? yes, it fired a few staples which is why they ended up having to resect more than expected because it had incomplete resections if yes, were the staples formed properly? yes, they were formed if yes, was the staple line complete? no.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.No lot or batch number was provided; therefore, a device history could not be done.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
It was reported that during an unspecified surgical procedure, the surgeon used pcee45a with a green load, fired the stapler, it sounded like it was going and then it sounded like the battery died.Surgeon said that it would not open with the reverse knife button so surgeon had to use the manual override.Surgeon opened a new stapler with green reload and it did the same thing.So surgeon then opened a pcee60a and it did it again.Surgeon then opened a ech60s and it worked.The surgeon said that the tissue tore while using the manual release function and that since the stapler did end up firing a few staples and cutting that surgeon had to resect more tissue.It added an additional 30 minutes to the procedure.There was no patient consequence reported.No additional information provided.
|
|
(b)(4).Date sent: 4/28/2026.D4: batch #920d12.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one pcee60a device was returned with no apparent damage and with no reload present.The device was tested for functionality in the straight position with a test reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples met the staple release criteria.The event described could not be confirmed as the device performed as intended and it opened and closed without any difficulties noted.The knife reverse button worked properly during testing.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 920d12, and no non-conformances were identified.
|