|
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the product returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis revealed that the distal connection tip of the cor/tri ant stem insert shaft was broken off, fragment was not returned for evaluation.The fracture is consistent with cyclic bending loading of the inserter tip that exceeded the fatigue strength of the material.Previous investigations based on failure analysis through complaint data, identified that exposure to high-cycle impact loading, particularly when combined with slightly off-center and/or irregular impact patterns, can induce cyclic bending stresses on the inserter tip.These conditions may significantly accelerate the propagation of fatigue cracks, ultimately leading to the mechanical failure of the impactor tip.Properly handling and attention to the approved use of the device diminishes the risk of failure.A dimensional inspection and functional testing were not performed as it is not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the cor/tri ant stem insert shaft would contribute to a device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history: a device history record (dhr) review or manufacturing records evaluation (mre) was not possible because the date code pg1108 provided is not a valid finished goods lot number.
|