| Catalog Number |
ECH45C |
| Medical Device Problem Codes |
Failure to Form Staple (2579); No Apparent Adverse Event (3189); Firing Problem (4011)
|
| Health Effect - Clinical Codes |
Pneumothorax (2012); No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an unknown procedure, it was observed that the knife moved forward for a moment, but then repeatedly stopped and moved forward, and the knife did not move forward easily at the first firing after opening the package and while transecting the bronchus.When reaching the tip of the jaw and returning, the knife returned the same movement, with repeatedly stopped and moved forward.After opening the jaw and checking the staples, the staples were malformed.The cartridge color was white.Another device was used to complete the case.There were no adverse consequences to the patient.No additional information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) date sent: 4/13/2026 b3: unknown d4: batch #unk d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.D4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.H4: the device manufacture date is currently not available.Additional information was requested, and the following was obtained: the procedure was thoracoscopic lobectomy.The staple formation style could not be clearly confirmed.At the time of the event, an air leak from the bronchus was observed.The area where staples could not be properly formed was additionally sutured with sutures.No change in the surgical procedure was required during additional suturing.As of the day after surgery, the patient¿s condition is stable.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.No lot or batch number was provided; therefore, a device history could not be done.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|