• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. CORFLO NASOGASTRIC/NASOINTESTINAL FEEDING TUBES; DH CPK NG TUBES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. CORFLO NASOGASTRIC/NASOINTESTINAL FEEDING TUBES; DH CPK NG TUBES Back to Search Results
Model Number 40-9438
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported via fda medsun uf/importer report (b)(4), the user facility reported that during a hypertensive episode, the patient experienced abdominal pain and imaging (kidney, ureter, bladder x-ray) revealed that the nasogastric tube was broken within the stomach.The tube had been in place for less than 30 days.During a gastrointestinal procedure, the nasogastric tube was identified as ruptured but not disconnected and was removed in its entirety under direct visualization.The esophageal and gastric mucosa were reported as within normal limits.A new nasogastric tube was placed and position was confirmed via imaging.The event occurred in february 2026.
 
Additional Manufacturer Narrative
The actual complaint product was not returned for evaluation.A review of the device history record and udi are in-progress.All information reasonably known as of 13 apr 2026 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORFLO NASOGASTRIC/NASOINTESTINAL FEEDING TUBES
Common Device Name
DH CPK NG TUBES
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT SA DE R.L. DE C.V. (AVENT 1)
circuito industial no.40
colonia obrera
nogales, cp
Manufacturer Contact
tamara cardona
5405 windward parkway
alpharetta, GA 30004
4704485487
MDR Report Key24865130
Report Number9611594-2026-00191
Device Sequence Number9239019
Product Code KNT
UDI-Device Identifier00350770460505
UDI-Public(01)00350770460505(240)104605002(17)301125(10)30385009
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K821906
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number40-9438
Device Catalogue Number40-9438
Device Lot Number30385009
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/16/2026
Supplement Date Received by Manufacturer04/30/2026
Initial Report FDA Received Date04/13/2026
Supplement Report FDA Received Date05/27/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/01/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age24 MO
Patient SexMale
-
-