| Catalog Number |
101025 |
| Medical Device Problem Codes |
Break (1069); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that an external fluid leak was observed from the "broken line" of a gambro cartridge set during priming.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: the actual device was not available; however, photographs of the sample were provided for evaluation.Visual inspection of the photos showed breakage on the arterial side of the cartridge, which allowed the reported leakage.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The reported condition was verified.The cause of the damage could not be determined as no further or actual sample testing could be conducted.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|