| Model Number |
1125564B |
| Medical Device Problem Code |
Circuit Failure (1089)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/12/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A device was returned to a service center in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.There was no report of patient harm or injury.A trilogy 200 ventilator was returned to a third-party service center for foam remediation on the device.There was no harm or injury reported.During the evaluation of the device at the third-party service center, there were no foam particles found in the device.The device failed functional testing relating to the dp purge valve test step.The service technician states that the trilogy 200/202/o2 sensor printed circuit assembly (pca) board was replaced to resolve the issue.The device passed final testing.
|
| |
|
Additional Manufacturer Narrative
|
|
The capa process identified complaints not being created for quality issues found during third party servicing.Based on a review of records impacted by the identified issue, it was determined that this record needed to be corrected and is now being submitted accordingly.
|
| |
|
Search Alerts/Recalls
|