| Model Number |
471344-19 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/29/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The curved bipolar dissector was analyzed and found to have thermal damage to the yaw pulley.The yaw pulley had charring and/or localized melting on the outer surface of one bipolar yaw pulley at the base of the grip.As a result of the damage, a section of the active electrode (grip) is exposed that would not normally be exposed.The instrument passed the electrical continuity test.The complaint was confirmed by failure analysis.The probable root cause of thermal damage is attributed to carbonized tissue on the grips of this bipolar instrument creating a conductive path during use.
|
| |
|
Event or Problem Description
|
|
It was reported that after a da vinci-assisted surgical procedure, the curved bipolar dissector instrument had surface damage.The procedure was completed with no reported injury.
|
| |
|