COVIDIEN SURGICAL TRI-STAPLE 2.0; STAPLE, IMPLANTABLE Back to Search Results
Model Number SIGTRSB60AMT
Medical Device Problem Codes Misfire (2532); No Apparent Adverse Event (3189)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/27/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the procedure, the reload was difficult to fire when used with a manual handle.The handle and reload were both replaced to resolve the issue.No patient complications have been reported as a result of this event.Medtronic's initial evaluation of the incident found that the reload was partially fired with the interlock engaged.
 
Additional Manufacturer Narrative
D10.Concomitant product: egiaustnd, egiaustnd endogia ultra univ std stap (lot # p5e1168).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Duplicate of (b)(4).
 
Additional Manufacturer Narrative
Additional information: b5, g3, h6 information received shows that this event is a duplicate of another issue reported under regulatory report#: 2647580-2026-01314.This file will be closed and all further updates processed under the above-referenced report for this event.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-STAPLE 2.0
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer (Section G)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key24867087
Report Number1219930-2026-02438
Device Sequence Number21470014
Product Code GDW
UDI-Device Identifier10884521717626
UDI-Public(01)10884521717626(17)270430(10)N4E2301Y(20)02
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K173270
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSIGTRSB60AMT
Device Catalogue NumberSIGTRSB60AMT
Device Lot NumberN4E2301Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/08/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/10/2026
Supplement Date Received by Manufacturer05/28/2026
Initial Report FDA Received Date04/13/2026
Supplement Report FDA Received Date06/09/2026
Date Device Manufactured05/31/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-