• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTCARE LTD F&P EVAQUA 2; BREATHING CIRCUIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTCARE LTD F&P EVAQUA 2; BREATHING CIRCUIT Back to Search Results
Model Number RT380
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/12/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Method:the subject mr290v autofeed humidification chamber provided as part of the rt380 adult dual heated evaqua2 breathing circuit and additional information, were requested to be provided to fisher & paykel (f&p) healthcare for evaluation.Results: there was no response to f&p healthcare's requests for additional information and return of the subject device.Conclusion: based on the limited information provided by the user and without the return of the subject device, we are unable to confirm or determine the cause of the reported issue.Every mr290v vented autofeed humidification chamber is pressure tested following the manufacturing process and any holes, cracks, or leaks would be identified during this process.In addition, the pressure test is following by a visual inspection of each chamber.Any chamber which fails these tests is rejected.The subject mr290v vented autofeed humidification chamber would have met the required specifications at the time of production.The user instructions for the rt380 adult dual heated evaqua2 breathing circuit state the following: - perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.- appropriate patient monitoring (e.G.Oxygen saturation) must be used at all times.- ensure appropriate ventilator or flow source alarms are set before connecting breathing sets to patient.- visually inspect breathing sets for damage (e.G.Crushed tube or cracked connector) before use and replace if damaged.- do not soak, wash, or sterilize this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers.- use usp sterile water for inhalation or equivalent for humidification.Do not add other substances to the water.
 
Event or Problem Description
A healthcare facility in china reported via a distributor that an mr290v vented autofeed humidification chamber provided as part of the rt380 adult dual heated evaqua2 breathing circuit was found leaking water during use.There was no reported patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
F&P EVAQUA 2
Common Device Name
BREATHING CIRCUIT
Manufacturer (Section D)
FISHER & PAYKEL HEALTCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurce paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
diego vargas banuelos
17400 laguna canyon road
suite 300
irvine, CA 92618
9494341005
MDR Report Key24867651
Report Number9611451-2026-01269
Device Sequence Number9100413
Product Code BZE
UDI-Device Identifier09420012429728
UDI-Public(01)09420012429728(10)2103915346(11)250820(17)300820
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K122432
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 04/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberRT380
Device Catalogue NumberRT380
Device Lot Number2103915346
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date04/13/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/20/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-