| Model Number |
31G X 6MM |
| Medical Device Problem Codes |
Material Twisted/Bent (2981); Fail-Safe Did Not Operate (4046)
|
| Health Effect - Clinical Code |
Needle Stick/Puncture (2462)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The complaint concern bent needle through the hole which caused accidental needle stick injury of nurse.No information about performed test or health consequences were provided.Htl-strefa made several attempt to obtain additional information to the event.The evaluation of archival samples and device history records (dhr) review did not confirm any defects.Product involved in incidents was not returned to htl-strefa s.A.For further evaluation.In the instruction for use there is below information in point 7.Insert the needle into the flat skin at a 90° angle in one continuous movement until the shield is fully recessed.Deliver the dose.Do not remove the needle until the dose has been completely delivered.Count to 10 before removing the needle from the skin to prevent drug leakage.Do not change the direction of the safety pen needle while it remains in the body as this can result in bending or breaking of the needle.We are aware that, the accidental injury might have led or might lead to the accidental blood exposure (abe).It is defined as an accident associated with exposure to blood, bloody fluids or other fluids, and might have caused or might cause serious infection in the injured third party.Due to fact that the event could potentially cause harm and taking into account that health care professional experience difficulties with using the device, we conclude this event should being reported.The final recommendation of hhe team is: to report the event in us where the event was occurred.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2026 htl-strefa inc.Received an email complaint notification.Customer stated: the needle came out of the side and stuck a nurse.We tested multiple ones and it happens with them all.Supposed to be a safety needle.Customer was called several times with no answer and no option for leaving a call back message.Email was sent for additional information.Sample under complaint has not been returned to htl-strefa until this report.
|
| |
|
Event or Problem Description
|
|
On 03/23/2026 htl-strefa inc.Received an email complaint notification.Customer stated: the needle came out of the side and stuck a nurse.We tested multiple ones and it happens with them all.Supposed to be a safety needle.Customer was called several times with no answer and no option for leaving a call back message.Email was sent for additional information.On 2026-04-22 sample under complaint has been returned for further analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint concern bent needle through the hole which caused accidental needle stick injury of nurse.No information about performed test or health consequences were provided.Htl-strefa made several attempt to obtain additional information to the event.The evaluation of archival sample, customers sample and device history records (dhr) review did not confirm any defects.Product involved in incidents was not returned to htl-strefa s.A.For further evaluation.In the instruction for use there is below information in point 7.Insert the needle into the flat skin at a 90° angle in one continuous movement until the shield is fully recessed.Deliver the dose.Do not remove the needle until the dose has been completely delivered.Count to 10 before removing the needle from the skin to prevent drug leakage.Do not change the direction of the safety pen needle while it remains in the body as this can result in bending or breaking of the needle.We are aware that, the accidental injury might have led or might lead to the accidental blood exposure (abe).It is defined as an accident associated with exposure to blood, bloody fluids or other fluids, and might have caused or might cause serious infection in the injured third party.Due to fact that the event could potentially cause harm and taking into account that health care professional experience difficulties with using the device, we conclude this event should being reported.
|
| |
|