| Model Number |
IPN920780 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
02/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that " an incident occurred on (b)(6) 2026.While inserting the epidural, the needle bent, and the anaesthetist stated that they had not applied excessive force.Device was changed, and there was a delay in pain management.".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The reported complaint of the needle was bent was confirmed based on the sample received.Visual examination of the returned sample revealed the needle appeared to be bent.The cannula was bent toward the center of the cannula.A device history record review was performed on the epidural needle with no relevant findings.It is unknown how the needle was handled prior to and during use and the pressure used for insertion and removal.Therefore, based on the information provided and the condition of the sample received, the potential cause of this complaint could not be determined.Further investigation has been initiated by the anesthesia research and development team within the teleflex quality system to determine a probable cause.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|
Event or Problem Description
|
|
It was reported that " an incident occurred on (b)(6) 2026.While inserting the epidural, the needle bent, and the anaesthetist stated that they had not applied excessive force.Device was changed, and there was a delay in pain management.".
|
| |
|