| Model Number |
NOXBOX-I |
| Medical Device Problem Code |
Gas Output Problem (1266)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/06/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
(e1) reported that while delivering inhaled nitric oxide (ino) therapy to an adult patient using the (d4) device, fluctuations in the monitored no concentration were observed.According to the hospital report, troubleshooting was performed; however, the fluctuations did not resolve.The device was subsequently removed from the patient and exchanged for another unit, after which the dosing issues were resolved.No harm to the patient or sustained changes in the patient's clinical status were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to the u.S.Service center for evaluation and underwent a comprehensive functional inspection and performance verification.All testing was performed in accordance with approved service and test procedures.During the evaluation, the internal manifold assembly used for gas-delivery control was identified as the cause of the issue.Based on the results of the device evaluation and the information available, the manifold was replaced.The device subsequently met all manufacturer specifications for nitric oxide delivery, gas monitoring, and all other functional performance checks.A review of complaint history for this failure mode indicated that the occurrence rate remains within established control limits.
|
| |
|