| Model Number |
414005 |
| Medical Device Problem Codes |
Fracture (1260); Ambient Noise Problem (2877)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/03/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Patient had what appeared to be true episodes of at/vt followed by a jump in rv short intervals and nst/vf episodes showing lead noise on all channels (b)(6) 2023.Last message on home monitoring was (b)(6) 2023.No impedance trends observed.The noise could be reproduced and a fracture is possible.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes biotronik submits this report to comply with fda regulations 21 cfr parts 4 and 803.Biotronik has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by biotronik, or its employees that the device, biotronik, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Biotronik will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes the lead is currently not available for analysis.No conclusion can be drawn based on available information at the moment.The file is closed.The investigation will be re-opened should additional data become available.
|
| |
|