| Model Number |
SET-0014-1 |
| Medical Device Problem Codes |
Leak/Splash (1354); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The following has been reported: customer reported: "it states that the blood tubing is leaking from the secondary site.Lot number fa25a21013.A preliminary review identified the following issue: administration set leak (external part) unknown if an active infusion was stopped.Reporting as a conservative measure.No adverse effects were reported.Additional information is needed to complete the investigation.
|
| |
|
Event or Problem Description
|
|
No sample or picture are available for analysis.Without the proper sample or picture evaluation is not possible to determine the probable root cause of the reported failure.The customer complaint cannot be confirmed.The most probable root cause of the reported incident is related to the lack of solvent application during the manufacturing process.The operator did not perform the joining operation correctly.Please note that without proper sample evaluation the exact root cause cannot be determined.The current process controls detection include 1) post sterilization sampling final inspection, 2) 100% visual inspection related to separated components during the manufacturing process and final physical inspections, 3) the first piece inspection will be performed to the first final assembled kit manufactured per shift.This kit will be submitted to visual inspection as per the classification of defects included in this specification and as per the applicable drawing.The batch record fa25a21013 was reviewed.No exceptions were generated that could classify as a possible root cause of this defect.The finished good lot has passed all sampling acceptance criteria for all tests performed including in-process testing and product testing.
|
| |
|
Search Alerts/Recalls
|