A device was returned to a service center in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.There was no report of patient harm or injury.A garbin ventilator was returned to a third-party service center for foam remediation on the device.There was no harm or injury reported.During the evaluation of the device at the third-party service center, the device was visually inspected and found no evidence of foam degradation.The foam was replaced preventatively.The device failed functional testing relating to the active exhalation purge valve closed test step.There is no documentation stated on a root cause or a component replacement.Run-in testing was performed again and the device passed.Additionally the device failed a not-reportable test step relating to check leak.The customer requested the return of the device unserviced, therefore no repairs were performed.
|