| Catalog Number |
SMS002 |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that prhc had two faulty rectal tube incidents.In the first incident, the dignishield balloon had a slow leak and fell out.In the second incident, the balloon would not deflate to remove from the patient.They had 3 remaining from this lot and had been instructed not to use due to patient safety.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because this investigation did not result in any additional findings and no sample was available for evaluation.Instructions for use were reviewed for this investigation.The labeling is found to be adequate, and it states, removal of device a.To remove the catheter from the rectum, the retention cuff must be deflated.1) attach the 50 ml syringe to the green inflation port and slowly withdraw all water from the retention cuff.B.Make sure the inflation port remains parallel to the catheter in order to prevent kinking in the tubing and blockage of fluid.Refer to ideal patient positioning described in step 3a.1) once you have ensured the cuff is fully deflated, disconnect the syringe and discard.2) to facilitate removal, you may add an appropriate amount of lubricant to the anal sphincter prior to pulling back on the catheter to remove the cuff.3) grasp the catheter as close to the patient as possible and slowly slide the cuff out of the anus.Dispose of the device in accordance with institutional protocol for disposal of medical waste.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.Corrections: d, h udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that prhc had two faulty rectal tube incidents.In the first incident, the dignishield balloon had a slow leak and fell out.In the second incident, the balloon would not deflate to remove from the patient.They had 3 remaining from this lot and had been instructed not to use due to patient safety.
|
| |
|