• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH SONATA; Cochlear Implant

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH SONATA; Cochlear Implant Back to Search Results
Model Number Mi1260 SONATA 2
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Hematoma (1884); Swelling/ Edema (4577)
Date of Event 01/01/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event or Problem Description
A revision surgery was performed to clean and drain a hematoma.
 
Event or Problem Description
The user, who is bilaterally implanted, presented two months ago with a spontaneous hematoma, without trauma, in the right temporal region over the magnet of the right cochlear implant.A revision surgery was performed to clean and drain the hematoma in (b)(6) 2026.During the procedure, it was observed that the magnet was outside the pocket.
 
Additional Manufacturer Narrative
Additional information: based on the received information from the field, the recipient underwent a surgery to drain a hematoma that has not been described in detail, at the implant site.During this procedure it was noticed that the implant magnet was out of the magnet pocket.However, it could be pushed back into the pocket and the surgery was completed successfully without further issues.The magnet very likely got out of the pocket during the surgical process by inadequate handling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SONATA
Common Device Name
Cochlear Implant
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
+43()577885626
MDR Report Key24882533
Report Number9710014-2026-00357
Device Sequence Number18604671
Product Code MCM
UDI-Device Identifier09008737385446
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/05/2026
Device Model NumberMi1260 SONATA 2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2026
Supplement Date Received by Manufacturer03/31/2026
Initial Report FDA Received Date04/15/2026
Supplement Report FDA Received Date09/15/2026
Date Device Manufactured02/05/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age3 YR
Patient SexMale
-
-