| Model Number |
Mi1260 SONATA 2 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Hematoma (1884); Swelling/ Edema (4577)
|
| Date of Event |
01/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
A revision surgery was performed to clean and drain a hematoma.
|
| |
|
Event or Problem Description
|
|
The user, who is bilaterally implanted, presented two months ago with a spontaneous hematoma, without trauma, in the right temporal region over the magnet of the right cochlear implant.A revision surgery was performed to clean and drain the hematoma in (b)(6) 2026.During the procedure, it was observed that the magnet was outside the pocket.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: based on the received information from the field, the recipient underwent a surgery to drain a hematoma that has not been described in detail, at the implant site.During this procedure it was noticed that the implant magnet was out of the magnet pocket.However, it could be pushed back into the pocket and the surgery was completed successfully without further issues.The magnet very likely got out of the pocket during the surgical process by inadequate handling.
|
| |
|