| Model Number |
3143 |
| Medical Device Problem Code |
Therapeutic or Diagnostic Output Failure (3023)
|
| Health Effect - Clinical Code |
Inadequate Pain Relief (2388)
|
| Date of Event |
03/18/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported the patient experienced ineffective stimulation.Surgical intervention may be pending to address the issue.Note : it is unknown which lead is related to this issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event is estimated.Additional components potentially involved in the event include: common device name: quattrode, model: 3143, udi: (b)(4), serial: n/a, batch: 117279.Common device name: octrode, model: 3186, udi: (b)(4), serial: na , batch: 119134.
|
| |
|
Event or Problem Description
|
|
Additional information was received wherein the lead was explanted and replaced.Effective therapy restored.
|
| |
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|