| Model Number |
MI1200 SYNCHRONY |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
Head Injury (1879)
|
| Date of Event |
04/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Following a head trauma, high impedances were reportedly observed on all channels.The user was reimplanted on (b)(6) 2026.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: the investigation results revealed overload fractures in the active electrode confirming that the device has failed due to an external impact to the active electrode.All the problems given in the recipient report appear to match well with this finding.This is a final report.
|
| |
|
Event or Problem Description
|
|
Following a head trauma, high impedances were reportedly observed on all channels.The user was reimplanted on (b)(6) 2026.
|
| |
|
Search Alerts/Recalls
|