| Model Number |
LF-GP |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A definitive root cause could be identified.Based on the results of the investigation, it is likely the following led to the malfunction: expected or random component failure without any design or manufacturing issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
The customer did not report a malfunction.During inspection of the fiberscope, glue was missing from the objective and light guide lens was found.There was no patient involvement.
|
| |
|