• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION DYNAGEN? MINI ICD VR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION DYNAGEN? MINI ICD VR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number D020
Medical Device Problem Codes High impedance (1291); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/31/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) was found to be exhibiting fault code 1005 prior to hip surgery.Technical services (ts) recommended system replacement.This icd remains in service.The patient is lost to follow up, so no additional information is available.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
This device remains in service and was not returned.As such, physical analysis has not been conducted in our laboratory.Based on the available information, and in the absence of a product return, the conclusion code cause not established was selected.As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.A risk review was completed and confirmed that the event of device displays error message was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance with labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.
 
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) was found to be exhibiting fault code 1005 prior to hip surgery.Technical services (ts) recommended system replacement.This icd remains in service.The patient is lost to follow up, so no additional information is available.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN? MINI ICD VR
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key24886945
Report Number2124215-2026-20364
Device Sequence Number18600617
Product Code LWS
UDI-Device Identifier00802526533860
UDI-Public(01)00802526533860(17)170825
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P960040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/25/2017
Device Model NumberD020
Device Catalogue NumberD020
Device Lot Number208703
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/17/2026
Supplement Date Received by Manufacturer05/11/2026
Initial Report FDA Received Date04/15/2026
Supplement Report FDA Received Date05/22/2026
Date Device Manufactured08/25/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age56 YR
Patient SexFemale
-
-