ALCON RESEARCH, LLC - HUNTINGTON CLAREON VIVITY TORIC EXTENDED VISION UV ABSORBING IOL; LENS, INTRAOCULAR, TORIC OPTICS
|
Back to Search Results |
|
| Model Number |
CCWET4 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The lens was returned for evaluation.Solution is dried on the lens.The lens is cut into five pieces.One haptic is broken in the gusset area and returned.Marks are observed on both sides of the optic that are similar to those left by a surgical instrument used to grasp the lens.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.The file indicates the use of a qualified cartridge with a non-qualified viscoelastic.Broken haptic damage was observed.The root cause is most likely a failure to follow the instructions for use (ifu).The viscoelastic indicated is not qualified for the lens/company cartridge combination used.Company foldable intraocular lenses (iols) are qualified for use with a company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that a company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.Company lens ifu: follow the section regarding directions for use for information on the maximum allowed time for the iol to stay in the folded condition.Failure to adhere to manufacturer¿s recommendations may result in iol damage.Ifu note: during lens loading and insertion, do not allow the company iol to remain in a folded condition within the selected iol delivery system for more than 3 minutes prior to completing insertion into the capsular bag.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A non-healthcare professional reported that during intraocular lens (iol) implantation procedure, when the lens was being inserted in left eye, the lens haptic broke off.The surgery was completed by removing the lens and replacing it with the same model and diopter company lens in an initial procedure.According to the additional information received, there was no patient harm.
|
| |
|
Search Alerts/Recalls
|
|
|