| Model Number |
DORNIER SMARTFLEX REUSABLE LIGHT GUIDE |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
It was reported that a fiber showed signs of burning and splitting at the blue tip.The dhr review confirmed that the fiber met all quality requirements prior to release.The fiber was not returned to dmta for evaluation and all other fibers from the suspect lot have been distributed.Past complaints were reviewed, and there is no identifiable trend in complaints related to holmium fibers burning.The risk associated with a fiber burning at the connector is identified as medium.Based on the available evidence, several factors may have contributed to this malfunction, including possible handling-related elements and other clinical use conditions.
|
| |
|
Event or Problem Description
|
|
Dmta received a complaint stating that a fiber showed signs of burning and splitting at the blue tip.
|
| |
|