| Model Number |
BF-P180 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the chipped cover malfunction: cause traced to degradation problem identified; expected or random component failure without any design or manufacturing issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
The bronchovideoscope was returned to the service center with a report of a suspicion of a leak.This does not indicate an mdr reportable event.However, mdr reportable failure was found during testing at the service center.During inspection of the device, a chipped distal end cover at the inner and outer channel opening was found.There was no patient involvement.
|
| |
|