| Catalog Number |
5C4482 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1: initial reporter phone no.: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a separation between the female connector and main body of the minicap transfer set.This occurred when disconnecting from a twin bag after peritoneal dialysis therapy.The transfer set was replaced because of the issue.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: the device was received for evaluation.Visual inspection was performed and a separation between the female connector and main body was observed.Leak, clear passage, and clamp function testing were performed, and no issues were observed.The reported condition was verified.The cause of the separation was determined to be manufacturing related due to an inadequate solvent bond between the female connector, insert chip, and main body.A nonconformance has been opened to address this issue.A batch review was conducted and there were no deviations found related to this condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|