| Model Number |
Mi1200 SYNCHRONY PIN |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The user reported a sudden loss of hearing with the device on 13-apr-2026.Reimplantation is considered.
|
| |
|
Event or Problem Description
|
|
The user reported a sudden loss of hearing with the device in (b)(6) 2026.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: based on the received information, a damage to the active electrode due to excessive mechanical stress seems very likely.However to determine an exact root cause device investigation would be necessary.Re-implantation is considered but no date has been scheduled yet.
|
| |
|
Event or Problem Description
|
|
The user reported a sudden loss of hearing with the device in (b)(6) 2026.The user was reimplanted.
|
| |
|
Event or Problem Description
|
|
The user reported a sudden loss of hearing with the device in (b)(6)2026.The user was reimplanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: based on the received information, a damage to the active electrode due to excessive mechanical stress seems very likely.However to determine an exact root cause device investigation would be necessary.The recipient has been reimplanted but the explanted device has not been received for investigation yet.This is a final report.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: overload fractures in the active electrode which are consistent with an external mechanical impact were determined to be the root cause of device failure.Other mechanical damages found are attributable to the removal surgery.The problems given in the recipient report appear to match the damage found.
|
| |
|