| Catalog Number |
00587806532 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
11/06/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).D10 - concomitant devices - persona porous standard femoral component size 9 right catalog #: 42502806602 lot #: 65715987, persona porous 2 peg tibial component size g right catalog #: 42530007902 lot #: 65588536, persona medial congruent vivacit-e articular surface right 10mm 8-11 gh catalog #: 42522100910 lot #: 65552692.Visual evaluation of pictures provided confirmed the patellar implant was fractured and exhibited signs of being explanted.Radiographic review additionally confirmed the patella was fractured with suspected loosening along the patellar post.The device history records were reviewed and no discrepancies were identified.A definitive root cause could not be determined with the information provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a right knee arthroplasty revision to address post-operative peg fracture of the patellar implant approximately seventeen (17) months post-operatively.Attempts have been made; however, no additional information is available.
|
| |
|