Additional information was added to d9, h3, h4, h6, and h11 h4: this lot was manufactured may 2025.H11: three (3) unopened samples were received for evaluation.Visual inspection and inspection under magnification were performed and revealed the following foreign matter enclosed in the sealed product packaging.Sample #1 0.2415 inch white liquid shape embedded in the transparent cap of white spike connector, inside the tubing, in the fluid pathway.Sample #2 0.4035 inch white liquid shape embedded in the transparent cap of white spike connector, inside the tubing, in the fluid pathway.Sample #3 0.1035 inch white liquid shape embedded in the transparent cap of white spike connector, inside the tubing, in the fluid pathway.The reported condition was verified.The cause of the condition was determined to be contamination during the manufacturing or packaging process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|