This complaint is recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the motor speeds were unstable, the machined head and width plates were damaged, and the reciprocating arm and screws were worn.The motor, machined head, pin, reciprocating arm, and screws were replaced to resolve the reported issue.The width plates were returned as is.Review of the device history record identified no deviations or anomalies during manufacturing.The root cause of the reported issue is attributed to component failure as the device exhibits wear and tear from repeated use.The reported event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.E1 telephone number: (b)(6).G2 country: (b)(6).
|