• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL STORZ SE & CO. KG ENDOFLATOR 50

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ SE & CO. KG ENDOFLATOR 50 Back to Search Results
Model Number UI500
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that error message were shown.During investigation error codes 247 and 382 are detected.No negative impact in state of health reported.
 
Additional Manufacturer Narrative
The device in question was returned to the manufacturer.This information is reflected in section d9.The evaluation is anticipated, but not yet begun.The investigation will be performed by a designated karl storz employee.The event is filed under internal karl storz complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOFLATOR 50
Common Device Name
ENDOFLATOR 50
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM  78532
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM   78532
Manufacturer Contact
anja fair
2151 e grand ave
el segundo, CA 90245
MDR Report Key24900083
Report Number9610617-2026-00849
Device Sequence Number18344517
Product Code FCX
UDI-Device Identifier04048551326657
UDI-Public04048551326657
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K161554
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberUI500
Device Catalogue NumberUI500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/01/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/01/2026
Initial Report FDA Received Date04/16/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-