| Model Number |
SPL-T |
| Medical Device Problem Code |
Corroded (1131)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for evaluation/inspection.Based on the results of the investigation, the rust was identified.It is likely the result of degradation; however, a definitive root cause could not be established.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the lithotripsy transducer exhibited rust on threaded end of the transducer where the ultrasonic probe is attached.There was no patient involvement.
|
| |
|