| Catalog Number |
S620GH/C |
| Medical Device Problem Codes |
Crack (1135); Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/02/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reports the unit is missing the liquid, apparently it has leaked or evaporated, leaving only the dry particles.No patient involved.A new device was used to complete the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
This event was identified during internal quality system activities.Upon review, the manufacturer determined the event meets fda criteria for reporting.The report is being submitted to ensure accuracy and completeness of mdr files.This submission does not reflect a change in device performance or patient risk.The awareness date corresponds to when the manufacturer became aware of the event details.All information reasonably known to the manufacturer at this time has been included.The suspect device was returned for evaluation.The device history record was reviewed, and no exception documents were found.A search of the complaint database was performed and no similar complaints for this lot number were found.This complaint will be used to trend for similar complaints.The complaint is confirmed; the luer lock is cracked.The root cause could not be determined.
|
| |
|