SISTEMAS MEDICOS ALARIS S.A DE C.V. ALARIS SMARTSITE; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2000E CHINA
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A 2000e china product was not available for investigation of pr (b)(4); however, the customer confirmed that the complaint sample was from lot 25025397.The feedback provided by the customer states that, " after connecting the infusion set to the needleless injection cap, the fluid failed to drip".No further information was available to assist the investigation in this instance.The details of this feedback were forwarded to the manufacturing site for investigation, where a review of the production records for lot 25025397 did not identify any in-process testing failures or quality deviations which may have caused or contributed to a report of this nature.The root cause of the customer¿s experience could not be determined in this instance as no sample was received for investigation.Without a sample to examine it is not possible to determine whether a manufacturing defect could have caused or contributed to a report of this nature.However, previous complaints for occlusions have been related to features on the surface of the male luer of the connecting products.These features include flash or a raised edge to the tip of the male luer which have previously been shown to intermittently lead to restricted flow due to them pinching the blue piston of the smartsite and not allowing it to open.This can sometimes be resolved by disconnecting and reattaching the same luer connection which may reposition the luer against the piston and improve the flow, or alternatively by changing the connecting male luer.
 
Event or Problem Description
It is reported, the fluid failed to drip.Despite repeated adjustments, the device could not be made to function properly, resulting in a delay in the patient¿s treatment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS SMARTSITE
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS S.A DE C.V.
parque sur, el realito
blvd. 2000
tijuana 22250
MX  22250
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS S.A DE C.V.
parque sur, el realito
blvd. 2000
tijuana 22250
MX   22250
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key24905184
Report Number9616066-2026-01002
Device Sequence Number9827445
Product Code FPA
UDI-Device Identifier07613203011938
UDI-Public(01)07613203011938
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K223088
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number2000E CHINA
Device Lot Number25025397
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/01/2026
Initial Report FDA Received Date04/16/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/19/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-