| Model Number |
L100 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/06/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that this device recorded a code 1003, indicating the voltage is too low for the projected remaining capacity.Data analysis identified potential high impedance within the battery.A device replacement timeline was created.The device will remain implanted and the patient will be monitored at this time.No adverse patient effects were reported.This device remains in service.
|
| |
|
Event or Problem Description
|
|
It was reported that this device recorded a code 1003, indicating the voltage is too low for the projected remaining capacity.Data analysis identified potential high impedance within the battery.A device replacement timeline was created.The device will remain implanted and the patient will be monitored at this time.No adverse patient effects were reported.This device remains in service additional information was received that this device was explanted and replaced.No additional adverse patient effects were reported.This device is no longer in service.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|