Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the product returned to depuy synthes for evaluation.Visual analysis revealed that zt trial sleeve exhibits an overall worn appearance consistent with normal and constant usage for a long period of time.Additionally, due to the worn condition of the device, the etching was impossible to read.The observed condition was identified as an end-of-life indicator, damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A dimensional inspection and functional testing were not performed as it is not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the zt trial sleeve would contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history: a manufacturing records evaluation (mre) was not performed as no lot number was retrieved for this device due to worn condition.
|