(b)(4).Date sent: 4/17/2026.D4 batch #: a9765x.Attempts are being made to obtain the following information.To date, no response has been provided.If further details are received at a later date, a supplemental medwatch will be sent: did the generator display any error messages and if so what were they? did the device pass the pre-test? had the device been working and then stopped? investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the device was returned with the distal tip of the blade broken off and it was returned with the device.The remaining blade portion was scratched and with evidence of contact with metal in or out of the operative field.During functional testing on gen11, an alert screen was displayed.A probable cause for the device to stop activating and the energy system to display an alert screen is blade damage.The device was analyzed, and it was determined that it was used in more than one generator.The device is intended to be used with a single generator.If the device is connected to a different generator an alert screen will be displayed indicating to replace instrument / no instrument uses remaining / restart generator.The device was disassembled to verify the condition of the internal components and no anomalies were found.Probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, staples, or clips during the procedure, or using any means other than the blade wrench to attach or detach the blade.Once minor blade damage has occurred, subsequent activations may increase the severity of the blade damage.This in turn can result in activation issues such as failing the pre-run test with the generator and displaying an alert screen.These alert screens that can result are such as ¿tighten assembly¿, ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure.Continued usage of the damaged blade can result in a broken blade.This device is packaged and sterilized for single use only.Multiple patient use may compromise the device integrity or create a risk of contamination that, in turn, may result in patient injury or illness.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot/batch a9765x, and no non-conformances were identified.
|