• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE FB TIB BASE SZ 8 CEM; ATTUNE IMPLANT : KNEE TIBIAL TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE FB TIB BASE SZ 8 CEM; ATTUNE IMPLANT : KNEE TIBIAL TRAY Back to Search Results
Catalog Number 150600008
Medical Device Problem Codes Loss of or Failure to Bond (1068); Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that patient was revised due to aseptic tibial loosening at cement to implant interface.Tibial implant was positioned with a slight anterior slope.Tibia debonded at the cement/implant interface.All components were removed.Femur was not loose.No delays or pt injury.Doi: (b)(6) 2014.Dor: (b)(6) 2026.Affected side: left knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Product description: attune fb tib base sz 8 cem.Product code: 150600008.Lot number: 7830037.Date of manufacturing: 01-jan-2014.Expiry date: 31-dec-2023.Quantity to be manufactured: (b)(4).A manufacturing record evaluation was performed for the finished device product description: attune fb tib base sz 8 cem, product code: 150600008, lot number: 7830037, and one non-conformances was identified, however not related to the reported failure mode of the complaint.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: product description- attune fb tib base sz 8 cem.Product code: 150600008.Lot number: 7830037.Date of manufacturing: 01-jan-2014.Expiry date: 31-dec-2023.Quantity to be manufactured: (b)(4).Added: d1, d10 concomitant.Corrected: h4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE FB TIB BASE SZ 8 CEM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key24911905
Report Number1818910-2026-07060
Device Sequence Number15728966
Product Code JWH
UDI-Device Identifier10603295042426
UDI-Public(01)10603295042426(17)231231(10)7830037
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K101433
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2023
Device Catalogue Number150600008
Device Lot Number7830037
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/01/2026
Supplement Date Received by Manufacturer05/25/2026
Initial Report FDA Received Date04/17/2026
Supplement Report FDA Received Date05/25/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/01/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE PS FB INSRT SZ 7 8MM; ATTUNE PS FEM LT SZ 7 CEM; SMARTSET GMV 40G US EO
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexMale
-
-