Model Number UM-S20-17S
Medical Device Problem Codes Material Deformation (2976); Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for inspection.Based on the results of the investigation, it is likely the following led to the malfunction: a degradation problem was identified regarding the kinks on the cable unit.A stress problem was identified regrading the physicl shaking of the probe the most probable cause is traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
It was observed that during the device evaluation, the ultrasonic probe exhibited kinks on probe¿s cable unit and the probe is physically shaking.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASONIC PROBE
Common Device Name
ULTRASONIC PROBE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key24912379
Report Number3002808148-2026-11985
Device Sequence Number18452050
Product Code ITX
UDI-Device Identifier04953170368479
UDI-Public04953170368479
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K250762
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberUM-S20-17S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2024
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/19/2026
Initial Report FDA Received Date04/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/12/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-