BECTON DICKINSON & CO (FRANKLIN LAKES) BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN 83 UNITS; TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION
|
Back to Search Results |
|
| Catalog Number |
367962 |
| Medical Device Problem Code |
Short Fill (1575)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There were multiple 510k numbers reported to be involved.The information is as follows: g.4.Pma / 510(k)#: k954592.Investigation summary: bd had not received any samples or photos for investigation.A total of 100 retained samples were visually inspected, with the hemogard closure assembly correctly assembled and placed on the tubes.Functional tests were performed on 10 retained samples.All tubes were within specification limits, with no issues identified.Based on a review of the device history record for the incident lots, all product specifications and requirements for lot release were met.This complaint has not been confirmed for the indicated failure mode: draw volume.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® pst¿ gel and lithium heparinn 83 units, an unspecified number of tubes underfilled.No adverse impact was reported regarding the patient or user.
|
| |
|