|
While no serious injury resulted in this event, if this malfunction recurred, it could cause or contribute to a serious injury or require medical or surgical intervention to preclude such.This event, therefore, is reportable per 21cfr part 803.Usb 3 firmware v1.4 contains an issue in the portion of code used to optimize sensor performance.If the optimization occurs during an exposure or sensor readout, there is a chance that the resulting image will be corrupted.Image data following the point of corruption will consist of data that was previously held in the memory of the usb device.As a result, the presented image may contain a portion of data from a previous exposure.In general, this would be obvious to the user since the presented image would either contain obvious artifacts or different anatomy than expected.
|