RESPIRONICS, INC. TRILOGY 200; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 1032800
Medical Device Problem Code Device Sensing Problem (2917)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/16/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The capa process identified complaints not being created for quality issues found during third party servicing.Based on a review of records impacted by the identified issue, it was determined that this record needed to be corrected and is now being submitted accordingly.
 
Event or Problem Description
A device was returned to a service center in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.There was no report of patient harm or injury.During the evaluation and functional testing of the device an error code related to o2 flow was identified.The device only came in for remediation, therefore no repair was performed.In addition, the device was visually inspected and found no evidence of foam degradation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 200
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key24913437
Report Number2518422-2026-112343
Device Sequence Number15934921
Product Code CBK
UDI-Device Identifier00606959022683
UDI-Public(01)00606959022683
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K093416
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation 505
Type of Report Initial
Report Date (Section B) 04/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number1032800
Device Catalogue Number1032800
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/16/2025
Initial Report FDA Received Date04/17/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/13/2010
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-