(b)(4).Date sent: 4/17/2026.D4: batch #657d22.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the pse45a device was returned with no apparent damage, and with no reload loaded on the device.Upon analysis, the jaws and closure trigger were noted in the close position, the knife was noted to be slightly advanced.The knife reverse switch was activated and the knife returned to the home position as expected.The device was tested for functionality with a test reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples met the staple form release criteria.The event described could not be confirmed as the device performed without any difficulties noted.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 657d22, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|