| Model Number |
72404252 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this inflatable penile prosthesis (ipp) implant device was non-functioning.A surgical procedure was performed during which the device was explanted and replaced.Upon explant, the physician identified fluid loss.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Model number/catalog number: 72404155.Serial number: n/a.Batch/lot number: 125731002.Model/catalog description: reservoir 65 ml pc/iz.Unique identifier (udi) #: (b)(4).The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.Review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.A risk review was completed and confirmed that the events of non-functional device and fluid leakage were defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
|
| |
|