| Catalog Number |
530.730 |
| Medical Device Problem Codes |
Mechanical Jam (2983); Complete Loss of Power (4015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11.Additional narrative: a visual and functional assessment was performed on the device.The following steps failed: check drill chuck function.During the service evaluation the following failures were identified: functional: frozen/will not move - chuck seized.Conclusion: failure reported is confirmed.Root cause: faulty parts.Action: repair.
|
| |
|
Event or Problem Description
|
|
It was reported that during service and evaluation, it was determined that the chuck with key f/bpl device was frozen/would not move - chuck seized.It was reported in the service order that the device was not working.It was unknown when the event occurred.There were no reports of delays to a surgical procedure.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
|
| |
|