DEXCOM, INC. DEXCOM G7 CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 9500-161
Medical Device Problem Code Wireless Communication Problem (3283)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that signal loss occurred.Performance data was reviewed.The allegation was confirmed due to the finding of signal loss over an hour within the investigation window.The probable cause was the transmitter and app were unable to establish a connection.No injury or medical intervention was reported.
 
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that signal loss occurred.Product was provided for investigation.An external visual inspection was performed and passed.Nordic bluetooth pairing test was performed and passed.Performance data was reviewed.The allegation was confirmed due to the finding of signal loss over an hour within the investigation window.The probable cause was the transmitter and app were unable to establish a connection.No injury or medical intervention was reported.
 
Additional Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G7 CONTINUOUS GLUCOSE MONITORING SYSTEM
Common Device Name
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
232 s dobson rd
mesa AZ 85202
Manufacturer Contact
jaime chavez
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key24918505
Report Number3004753838-2026-134031
Device Sequence Number10564998
Product Code QBJ
UDI-Device Identifier00386270004550
UDI-Public(01)00386270004550(11)251001(17)270331(10)1725300001
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K213919
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number9500-161
Device Catalogue NumberSTP-AT-018
Device Lot Number1725300001
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/25/2026
Supplement Date Received by Manufacturer05/26/2026
Initial Report FDA Received Date04/17/2026
Supplement Report FDA Received Date06/08/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/01/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age82 YR
Patient SexFemale
Patient Weight93 KG
-
-