| Catalog Number |
RED68KIT |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation of the returned penumbra system red 68 reperfusion catheter (red68) confirmed it was fractured and revealed yield marks near the fracture location with no signs of torquing or stretching.This indicates the fracture was likely due to a kink.If the red68 is manipulated against resistance, damage such as a kink and subsequent fracture may occur.Based on the reported complaint, the root cause of resistance could not be determined.Further evaluation revealed kinks on the catheter.This damage was incidental to the reported complaint, and the root cause could not be determined.Penumbra devices are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a thrombectomy procedure using a penumbra system red 68 reperfusion catheter (red68), a neuron max 6f 088 long sheath (neuron max), and a non-penumbra microcatheter.One pass was completed using the red68.After removing the red68 from the patient, the red68 was noticed to be fractured near the midshaft.There was no portion of the red68 that remained inside the patient.The procedure was completed at this point.There was no report of an adverse effect to the patient.
|
| |
|