Catalog Number 71364009
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during a tha surgery, upon final impaction of femoral head, the black plastic of the tip of a femoral head impactor broke into three pieces.The procedure was resumed, without any delay, using the same device.No further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).This complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Additional Manufacturer Narrative
H3, h6: the associated device was returned and evaluated.The visual inspection revealed that the device returned with wear from use evident, the laser markings are legible and the threads in the sleeve undamaged.The tip was not returned for evaluation.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history based on the historical data revealed a similar event for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL HEAD IMPACTOR
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key24919502
Report Number1020279-2026-00528
Device Sequence Number9897858
Product Code HRY
UDI-Device Identifier03596010199256
UDI-Public3596010199256
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K121393
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71364009
Device Lot Number03DM03205
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/01/2026
Supplement Date Received by Manufacturer04/23/2026
05/15/2026
Initial Report FDA Received Date04/18/2026
Supplement Report FDA Received Date05/18/2026
05/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-