ARTHROCARE CORPORATION FIRSTPASS MINI RIGHT; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Catalog Number 72290130
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/28/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a knee arthroscopy, a portion of the tip of the firstpass became loose.The procedure was resumed, without any delay, using a similar s+n back-up product.No further complications were reported.
 
Additional Manufacturer Narrative
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alter the conclusions of this report, a follow-up report will be submitted as required.
 
Additional Manufacturer Narrative
H11: internal complaint reference: (b)(4).B5 and h6 updated.
 
Event or Problem Description
It was reported that during a knee arthroscopy, a portion of the tip of the firstpass became loose.The pieces were removed from inside the patient using tweezers.The procedure was resumed, without any delay, using a similar s+n back-up product.No further complications were reported.
 
Additional Manufacturer Narrative
H11: internal complaint reference (b)(4).H3, h6: the reported device was not returned to the designated complaint unit for independent evaluation.However, an image evaluation was performed and found the firstpass mini device on a surface, the suture capture was detached from the upper jaw.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause could not be determined since findings cannot lead to a clear cause of the reported event.Factors that could have contributed to the reported event include: (1) excessive force (2) tissue thickness (3) damage or debris on the device tip between passes.Based on this investigation, the need for corrective action is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIRSTPASS MINI RIGHT
Common Device Name
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
ARTHROCARE CORPORATION
7000 west william cannon drive
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORPORATION
7000 west william cannon drive
austin TX 78735
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key24919767
Report Number3006524618-2026-00193
Device Sequence Number10494581
Product Code NBH
UDI-Device Identifier00885556694572
UDI-Public885556694572
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72290130
Device Lot Number2192770
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/28/2026
Supplement Date Received by Manufacturer05/04/2026
05/28/2026
Initial Report FDA Received Date04/18/2026
Supplement Report FDA Received Date05/04/2026
05/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/08/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age25 YR
Patient SexMale
-
-