SMITH & NEPHEW, INC. HEALICOIL KNOTLESS PK ST; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Catalog Number |
72205138 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/31/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that, during an arthroscopic rotator cuff repair, after the insertion site was prepared using a 5.5 mm healicoil tap and cleared of all debris, when a healicoil knotless anchor was inserted into the hole and the peek outer core was advanced into the hole, the peek coil broke.All the pieces as well as the metal tip of the anchor were retrieved from the surgical site using an arthroscopic grasper.The procedure was resumed, after a non-significant delay, using a similar s+n back-up, in the originally drilled bone hole.Additional anesthesia was required as consequence of the surgical prolongation.No further complications were reported.
|
| |
|
Search Alerts/Recalls
|
|
|